FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2062201 · Received April 11, 2011

Report

Report Number
1644487-2011-00766
Event Type
Malfunction
Date Received
April 11, 2011
Date of Event
January 1, 2011
Report Date
March 14, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ATTEMPT TO UPGRADE A PHYSICIAN'S DELL X50 HANDHELD DEVICE, A COMPANY REP NOTED THAT THE HANDHELD WILL NOT STAY ON . THE REP OBSERVED THAT THE POWER LIGHT WILL TURN OFF WHEN THE HANDHELD IS MOVED. IF THE SERIAL ADAPTER CABLE AND POWER CORD ARE MOVED AROUND, THE LIGHT WILL COME ON BUT WILL NOT STAY ILLUMINATED. THE PHYSICIAN'S OFFICE CONFIRMED THAT THIS ISSUE HAS BEEN OCCURRING FOR SEVERAL MONTHS AND THEY WERE USING A DIFFERENT HANDHELD ON THEIR VNS PATIENTS. WHEN THE COMPANY REP USED HIS OWN POWER CORD AND SERIAL ADAPTER CABLE, HE WAS ABLE TO USE THE HANDHELD WITHOUT ANY ISSUES. THE SITE HAS BEEN SENT REPLACEMENT CABLES/CORDS AND GOOD FAITH ATTEMPTS TO HAVE THE POWER CORD AND SERIAL ADAPTER CABLE RETURNED TO THE MFR FOR ANALYSIS HAVE BEEN UNSUCCESSFUL TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. 250 834042

Patients

Seq Age Sex Outcome Treatment
1