FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 2062059 · Received April 6, 2011

Report

Report Number
1720753-2011-03279
Event Type
Malfunction
Date Received
April 6, 2011
Date of Event
March 18, 2011
Report Date
April 6, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MFR'S SERVICE REP PERFORMED AN ON-SITE INVESTIGATION. THE SERVICE REP REPLACED THE SRAM. THE SYSTEM WAS TESTED AND OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE 9600 SYSTEM DISPLAYED A SOFTWARE CORRUPTION ERROR MESSAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1