FDA Adverse Event Malfunction Summary report: N

DARIO BLOOD GLUCOSE MONITORING SYSTEM

MDR report key: 20620190 · Received November 6, 2024

Report

Report Number
3010606081-2024-00064
Event Type
Malfunction
Date Received
November 6, 2024
Date of Event
October 6, 2024
Report Date
November 6, 2024
Manufacturer
LABSTYLE INNOVATIONS LTD.
Product Code
NBW
UDI-DI
00850065006154
PMA / PMN Number
K150817
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

WHILE ON THE PHONE WITH DARIO'S REPRESENTATIVE, IT WAS DETERMINED THAT THE USER WAS USING A CARTRIDGE OF STRIPS THAT WAS OPEN FOR MORE THAN 30 DAYS AND WAS THEREFORE EXPIRED. DARIO'S USER GUIDE STATES IN THE SECTION PRECAUTIONS: "DO NOT USE A TEST STRIP EXPIRED 1 MONTH FROM OPENING. DISCARD THE CARTRIDGE 1 MONTH AFTER OPENING". A REPLACEMENT OF DARIO'S STRIPS AND CONTROL SOLUTION WAS SENT TO THE USER TO MAKE SURE THAT THE DARIO STRIPS ARE READING CORRECTLY, AND TO ENSURE PROPER TECHNIQUE. DURING TROUBLESHOOTING ON THE PHONE WITH DARIO'S REPRESENTATIVE, THE USER MENTIONED THAT AFTER HE HAD TESTED WITH HIS NON-DARIO METER AND RECEIVED A BG READING OF 102 MG/DL HE THEN TESTED HIS BG WITH HIS SECOND DARIO METER, AND ALSO RECEIVED A READING OF 102 MG/DL. DUE TO THE DIFFERENCE IN READINGS ABOVE, A REPLACEMENT OF DARIO'S METER WAS SENT TO THE USER TOGETHER WITH A RETURN MATERIAL AUTHORIZATION (RMA) PACKAGE, TO FURTHER INVESTIGATE THIS ISSUE. THE RMA PACKAGE WAS RECEIVED FOR INVESTIGATION ON OCTOBER 30TH, 2024. THE METER WAS TESTED AND WAS READING WITHIN RANGE. RMA INVESTIGATION TEST RESULTS WITH DARIO'S CONTROL SOLUTION CONCLUDED THAT THE READINGS WERE WITHIN RANGE: LEVEL 1 TEST RESULTS WERE 133 MG/DL, 133 MG/DL, 133 MG/DL, WITHIN THE RANGE OF 109-157 MG/DL. LEVEL 2 TEST RESULTS WERE 223 MG/DL, 226 MG/DL, 227 MG/DL, WITHIN THE RANGE OF 182-261 MG/DL. THEREFORE, IT WAS DETERMINED THAT THIS COMPLAINT IS NOT DUE TO A METER ISSUE. AFTER THE NEW REPLACEMENT OF DARIO'S METER WAS RECEIVED BY THE USER, AN ATTEMPT WAS MADE TO FOLLOW UP WITH THE USER IN ORDER TO TEST THE USER'S BG READINGS WITH THE NEW DARIO METER, CARTRIDGE OF STRIPS AND CONTROL SOLUTION, HOWEVER, NO RESPONSE HAS BEEN RECEIVED TO DATE. THE HIGH BG READINGS MAY HAVE BEEN DUE TO MISUSE OF THE STRIPS. THERE IS NOT ENOUGH INFORMATION REGARDING THE OLD STRIPS AVAILABLE TO FURTHER INVESTIGATE THIS CASE. NO RESOLUTION IS AVAILABLE.

Description of Event or Problem · 0

ON OCTOBER 6TH, THE USER CONTACTED DARIO TO REPORT HIGH BLOOD GLUCOSE (BG) READINGS. THE USER REPORTED THAT ON THE MORNING OF OCTOBER 6TH, HE RECEIVED A BG READING OF 250 MG/DL FROM HIS DARIO METER. THE USER STATED THAT HE RETESTED HIS BG LEVEL 45 MINUTES LATER AND RECEIVED A BG READING OF 393 MG/DL. DUE TO THE ABOVE THE USER TESTED HIS BG LEVEL WITH A NON-DARIO METER AND RECEIVED A READING OF 102 MG/D, WHICH THE USER STATED WAS IN NORMAL RANGE FOR HIM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1997496 DARIO BLOOD GLUCOSE MONITORING SYSTEM GLUCOMETER NBW LABSTYLE INNOVATIONS LTD. 1021-04 2204211 00850065006154

Patients

Seq Age Sex Outcome Treatment
1 NA Male Other