FDA Adverse Event Malfunction Summary report: N

ALINITY I ANTI-HBS REAGENT KIT

MDR report key: 20619918 · Received November 6, 2024

Report

Report Number
3008344661-2024-00133
Event Type
Malfunction
Date Received
November 6, 2024
Date of Event
October 28, 2024
Report Date
December 19, 2024
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
LOM
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.  THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 7P89 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 7P88, WITH 510K/PMA/BLA NUMBER P050051.

Additional Manufacturer Narrative · 0

THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND FIELD DATA REVIEW. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. A REVIEW OF TRACKING AND TRENDING DID NOT IDENTIFY ANY TRENDS FOR THE ISSUE FOR THE PRODUCT. THE TICKET SEARCH BY LOT INDICATES THAT THE COMPLAINT LOT PERFORMS AS EXPECTED FOR THIS PRODUCT. DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES, POTENTIAL NON-CONFORMANCES, OR DEVIATIONS ASSOCIATED WITH THE COMPLAINT LOT AND COMPLAINT ISSUE. THE OVERALL PERFORMANCE OF THE ALINITY I ANTI-HBS REAGENTS WAS REVIEWED USING FIELD DATA GATHERED FROM CUSTOMERS WORLDWIDE. THE MEDIAN VALUES FOR COMPLAINT LOT 65163FZ00 ARE WITHIN THE ESTABLISHED LIMITS AND COMPARABLE TO THE HISTORICAL REAGENT LOT PERFORMANCE, WHICH CONFIRMS NO SYSTEMIC ISSUE FOR THE LOT. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER¿S ISSUE. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ALINITY I ANTI-HBS REAGENT, LOT 65163FZ00, WAS IDENTIFIED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY I ANTI-HBS RESULTS GENERATED FOR PATIENT SAMPLES. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT = 57, REPEAT = <10. INITIAL RESULT = 11.5, REPEAT = <10. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY I ANTI-HBS RESULTS GENERATED FOR PATIENT SAMPLES. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT = 57, REPEAT = <10. INITIAL RESULT = 11.5, REPEAT = <10 . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1416485 ALINITY I ANTI-HBS REAGENT KIT TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM) LOM ABBOTT IRELAND DIAGNOSTICS DIVISION 65163FZ00

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I PROCESSING MODU, 03R65-01, (B)(6)| ALNTY I PROCESSING MODU, 03R65-01, (B)(6)