REFURBED COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNE
Report
- Report Number
- 6000001-2011-02958
- Event Type
- Malfunction
- Date Received
- April 19, 2011
- Date of Event
- March 9, 2011
- Report Date
- March 10, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-3/15/05-007-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE 810:11 WAS FOUND IN THE PUMP'S EVENT HISTORY BUT NOT DUPLICATED DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE WAS AN AIR IN LINE BOARD OUT OF CALIBRATION. THEREFORE, THE PUMP HEAD MODULE WAS REPLACED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED FAILURE CODE 810:11 UPON POWER UP. DURING DEVICE EVALUATION, FAILURE CODE 810:11 WAS CONFIRMED AND INTERRUPTED DELIVERY. THERE WAS NO PATIENT INVOLVEMENT. THE SOFTWARE VERSION OF THIS DEVICE IS CURRENTLY UNKNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REFURBED COLLEAGUE CX VOLUMETRIC INFUSION PUMP SINGLE CHANNE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |