FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 2061911 · Received March 18, 2011

Report

Report Number
1824206-2011-01703
Event Type
Malfunction
Date Received
March 18, 2011
Date of Event
February 23, 2011
Report Date
February 23, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THE CPR MECHANISM HAD GREASE ON IT THAT HAD BECOME GUMMY. HE CLEANED ALL OF THE OLD GREASE FROM THE CPR MECHANISM AND RE-LUBRICATED THE ASSEMBLY TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE CPR FUNCTION WAS NOT FUNCTIONING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1