FDA Adverse Event Malfunction Summary report: N

6461500 INFUSOR PUMP (1 LABEL)

MDR report key: 2061902 · Received April 19, 2011

Report

Report Number
6000001-2011-02955
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
April 13, 2011
Report Date
April 13, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K883577
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE CONDITION OF AN INFUSO.R. THAT "WENT INTO TRIPLE ALARM AFTER DELIVERING 0.25 MILLILITERS DURING DEVICE EVALUATION, INTERRUPTING DELIVERY", WHICH WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE REED SWITCH OUT OF ADJUSTMENT. THE REED SWITCH WAS ADJUSTED IN ORDER TO FIX THE REPORTED CONDITION. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THIS PUMP HAS NEVER PREVIOUSLY BEEN SENT TO BAXTER FOR SERVICE AND, THEREFORE, THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS REPORTED PROBLEM FOR THIS PUMP. A DEVICE HISTORY REVIEW WAS ALSO PERFORMED, FINDING THAT NO EXCEPTIONS WERE NOTED DURING THE MANUFACTURING OF THIS DEVICE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THE BAXTER PRODUCT ANALYSIS LABORATORY TECHNICIAN REPORTED AN INFUSOR DEVICE WHICH WENT INTO TRIPLE ALARM AFTER DELIVERING 0.25 MILLILITERS DURING DEVICE EVALUATION, INTERRUPTING DELIVERY. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 6461500 INFUSOR PUMP (1 LABEL) PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1