FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2061854 · Received April 4, 2011

Report

Report Number
1644487-2011-00707
Event Type
Malfunction
Date Received
April 4, 2011
Date of Event
May 12, 2005
Report Date
March 14, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF PROGRAMMING HISTORY.

Description of Event or Problem · 1

DURING MFR REVIEW OF A PT'S VNS PROGRAMMING HISTORY, IT WAS NOTED THAT AN INTERRUPTED SYSTEMS TEST OCCURRED ON (B)(6) 2005 WHICH CAUSED THE SETTINGS TO CHANGE. THE CHANGE WAS IMMEDIATELY NOTED UPON RE-INTERROGATION AND ALL SETTINGS WERE CORRECTED EXCEPT FOR THE OFF TIME, WHICH REMAINED AT 60 MINS. THE OFF TIME WAS NOT CORRECTED UNTIL (B)(6) 2005.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. 250

Patients

Seq Age Sex Outcome Treatment
1 17 YR