FDA Adverse Event
Malfunction
Summary report: N
FORMULA TOM CAT CUTTER
MDR report key: 2061852
·
Received April 14, 2011
Report
- Report Number
- 2061852
- Event Type
- Malfunction
- Date Received
- April 14, 2011
- Date of Event
- April 12, 2011
- Report Date
- April 14, 2011
- Manufacturer
- STRYKER CORP
- Product Code
- FZT
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- FL, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
METAL FRAGMENTS WERE NOTICED IN THE JOINT SPACE/SHOULDER AREA. PICTURES ARE AVAILABLE.======================MANUFACTURER RESPONSE FOR FORMULA TOMCAT CUTTER, FORMULA TOM CAT CUTTER======================I TOLD ASCENT THAT THE PACKAGING WAS NOT SAVED AND REFERENCED FROM ANOTHER PRODUCT IN STOCK. WE WILL KEEP PRODUCT UNTIL WE ARE CONFIDENT THAT THE PATIENT DOES NOT DEVELOP AN INFECTION OR OTHER COMPLICATION.======== REPORTER'S RESPONSE ======================ACCORDING TO THE REPORTER, THIS IS THE FIRST INCIDENT THEY HAVE WITH THE DEVICE. NO OBSERVABLE MANUFACTURING PROBLEM NOTED. THE METAL FRAGMENTS WERE NOT REMOVED FROM THE PATIENT. "AFTER IRRIGATION THE FRAGMENTS WERE EMBEDDED INTO THE TISSUE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FORMULA TOM CAT CUTTER | FORMULA TOMCAT CUTTER | FZT | STRYKER CORP | 375-565-000 | 1116966-FROM OTHERS IN STOCK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |