FDA Adverse Event Malfunction Summary report: N

HEMOCRON

MDR report key: 2061802 · Received April 11, 2011

Report

Report Number
2061802
Event Type
Malfunction
Date Received
April 11, 2011
Date of Event
April 1, 2011
Report Date
April 11, 2011
Manufacturer
INTERNATIONAL TECHNIDYNE CORP.
Product Code
JPA
Product Problem
Yes
Report Source
User Facility report
Reporter Location
FL, US

Narratives

Description of Event or Problem · 1

WE WERE DOING A VASCULAR CASE WITH THE DOCTOR AND NEEDED TO USE THE HEMOCHRON MACHINE. THE RESULTS CAME BACK UNUSUALLY HIGH ACCORDING TO THE DOCTOR. HE THEN ASKED US TO RETEST. THE RESULTS ALSO CAME BACK HIGH. THEN THE DOCTOR ASKED US IF THE MACHINE HAD BEEN CALIBRATED THAT DAY. WE VERIFIED WITH THE CHARGE NURSE THAT IT HAD BEEN CALIBRATED THAT DAY. WE BORROWED ANOTHER HEMOCHRON FROM ANOTHER DEPARTMENT. THE SECOND MACHINE OPERATED WITHIN NORMAL PARAMETERS AND WE WERE ABLE TO PROCEED WITH OUT INCIDENT. THE MALFUNCTIONING HEMOCHRON MACHINE WAS IMMEDIATELY REMOVED FROM SERVICE AND SENT TO BIOMED FOR REPAIR.BIOMED COMMENTS: I DID CHECK THE HEMOCHRON UNIT ACCORDINGLY WITH THE MANUFACTURER INSTRUCTIONS. THE ONLY TWO WAYS TO TEST THE UNIT IT IS USING THE EQC AND THE LIQUID QC THAT IT IS SUPPOSED TO BE DONE BY THE POC ON THEIR OWN DAILY OR WEEKLY SCHEDULE. THE UNIT PASSED THE EQC THAT I DID. THIS EQC WILL TEST ALL THE MECHANICAL AND ELECTRICAL CIRCUITS IN THE UNIT. THE UNIT WAS PUT BACK INTO OPERATION. NO HARM TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEMOCRON HEMOCRON JPA INTERNATIONAL TECHNIDYNE CORP. * *

Patients

Seq Age Sex Outcome Treatment
1 85 YR