FDA Adverse Event Injury Summary report: N

SCREW, CANN. DELTA TAPERED BIO-INT. 10MM

MDR report key: 2061775 · Received April 19, 2011

Report

Report Number
1220246-2011-00070
Event Type
Injury
Date Received
April 19, 2011
Date of Event
December 22, 2010
Report Date
April 12, 2011
Manufacturer
ARTHREX, INC.
Product Code
MAI
PMA / PMN Number
K071176
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PATIENT DEMOGRAPHICS (AGE AT TIME OF EVENT, DATE OF BIRTH, GENDER, WEIGHT) WERE REQUESTED, BUT NOT PROVIDED. NO FURTHER PATIENT INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT OR MADE AVAILABLE IN RESPONSE TO FOLLOW-UP COMMUNICATION. NO ADDITIONAL ADVERSE CONSEQUENCES HAVE BEEN REPORTED FROM THIS EVENT. THIS DEVICE IS USED FOR TREATMENT. THE DEVICE WAS REQUESTED FOR EVALUATION BUT WAS NOT RETURNED, THEREFORE, THE COMPLAINANT'S EVENT COULD NOT BE VERIFIED. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. LOT NUMBER WAS REQUESTED BUT NOT PROVIDED, THEREFORE DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED. THE MOST LIKELY CAUSE(S) OF THIS TYPE OF EVENT IS DAMAGE TO THE IMPLANT AT THE TIME OF INSERTION WHICH LED TO THE BREAKAGE AND POST-OP FRAGMENT IN THE JOINT SPACE. THE POTENTIAL CAUSES OF THIS EVENT ARE BEING COMMUNICATED TO THE EVENT REPORTER. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED. UNKNOWN DEVICE DISPOSITION.

Description of Event or Problem · 1

THE SCREW POSSIBLY BROKE DURING AN OPERATION IN THE RIGHT KNEE. THE TIP OF THE SCREW HAD MIGRATED INTO THE KNEE JOINT AND HAD CAUSED A CARTILAGE INJURY. THIS WAS DISCOVERED DURING A REVISION SURGERY AND THE BROKEN SCREW WAS REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SCREW, CANN. DELTA TAPERED BIO-INT. 10MM FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE MAI ARTHREX, INC. UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other