VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2011-00253
- Event Type
- Malfunction
- Date Received
- April 19, 2011
- Date of Event
- March 24, 2011
- Report Date
- March 24, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE IS EN ROUTE TO THE MANUFACTURER. WE WILL PROVIDE A FINAL REPORT UPON RECEIPT OF THE DEVICE AND COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE RETURNED COMPLAINT CHAMBER WAS ATTACHED TO A WATERBAG FOR TESTING. RESULTS: THE CHAMBERS FILLED TO THE EXPECTED FILL LEVEL AND THEN SMALL DROPS OF WATER BEGAN TO LEAK FROM THE CONNECTION BETWEEN THE TUBE AND THE WATERBAG SPIKE. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR THIS LOT NUMBER. CONCLUSION: ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE LEAK DEVELOPED POST PRODUCTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT". (B)(4).
A HOSPITAL IN (B)(6) REPORTED THAT WATER LEAKED FROM THE CONNECTION PART BETWEEN THE FEEDSET TUBE AND THE BAG SPIKE OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER AFTER TWELVE DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.
A HOSPITAL IN (B)(6) REPORTED THAT WATER LEAKED FROM THE CONNECTION PART BETWEEN THE FEEDSET TUBE AND THE BAG SPIKE OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER AFTER TWELVE DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 100319 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |