FDA Adverse Event Malfunction Summary report: N

BUBBLE CPAP CIRCUIT

MDR report key: 2061712 · Received April 19, 2011

Report

Report Number
9611451-2011-00248
Event Type
Malfunction
Date Received
April 19, 2011
Report Date
March 23, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZD
PMA / PMN Number
REFER TO H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE BC151 IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K100011. METHOD: THE RETURNED BC151 BUBBLE CPAP CIRCUIT KIT WAS VISUALLY INSPECTED FOR A MISSING CONNECTOR/ADAPTER. RESULTS: VISUAL INSPECTION REVEALED THAT THE CONNECTOR/ADAPTER WAS NOT MISSING BUT RATHER AN INCORRECT TYPE OF INSPIRATORY TUBE HAS BEEN PACKED WITH THE CIRCUIT KIT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 110126. CONCLUSION: THE BUBBLE CPAP CIRCUIT KIT CONSISTS OF A NUMBER OF COMPONENTS GROUPED TOGETHER DURING PRODUCTION. IT IS MOST LIKELY THAT OPERATOR ERROR HAS RESULTED IN AN INCORRECT TYPE OF INSPIRATORY TUBE HAVING BEEN PACKED WITH THIS BUBBLE CPAP CIRCUIT KIT. THIS IS MOST LIKELY DUE TO A FAULT IN THE LINE CLEARANCE PROCESS. THE NEW STANDARD OPERATING PROCEDURES HAVE BEEN PUT IN PLACE TO ASSIST THE OPERATORS ON THE PRODUCTION LINE TO CORRECTLY PACK THE CIRCUIT KITS. A SPECIFIC PACK CARD DETAILING ALL COMPONENTS REQUIRED MUST BE DISPLAYED AT THE PACKING STATION. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT A CONNECTOR/ADAPTER WAS MISSING FROM A BC151 BUBBLE CPAP CIRCUIT KIT. THIS WAS NOTICED PRIOR TO PATIENT USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BUBBLE CPAP CIRCUIT BZD BZD FISHER & PAYKEL HEALTHCARE LIMITED BC151 110126