ETS FLEX
Report
- Report Number
- 3005075853-2011-01557
- Event Type
- Malfunction
- Date Received
- April 18, 2011
- Date of Event
- March 10, 2011
- Report Date
- March 29, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD PRESENT. THE RELOAD WAS RECEIVED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED AND CUT WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE.
IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, WHEN THE DEVICE WAS FIRED ON THE INFERIOR MESENTERIC ARTERY, THE GUN WAS LOADED WITH A WHITE RELOAD AND WAS FIRED FOR THE FIRST TIME. HOWEVER THE STAPLES FAILED TO FORM CORRECTLY AND DID NOT FORM A COMPLETE STAPLE LINE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETS FLEX | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RELOAD PRODUCT #-TR45W |