FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES

MDR report key: 20616996 · Received November 6, 2024

Report

Report Number
9617032-2024-01740
Event Type
Malfunction
Date Received
November 6, 2024
Date of Event
October 7, 2024
Report Date
December 5, 2024
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
30382903673767
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES, BUT 5 PHOTOS WERE PROVIDED FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR SAMPLE QUALITY- FIBRIN WAS OBSERVED. ADDITIONALLY, 10 RETENTION SAMPLES OF EACH LOT FROM BD INVENTORY WERE EVALUATED BY FUNCTIONAL TESTING AND NO ISSUES WERE OBSERVED AS ALL SAMPLES MET SPECIFICATIONS. COMPLAINTS FOR SAMPLE QUALITY ARE UNDER STATISTICAL CONTROL FOR THE MONTH OF OCTOBER 2024. AT THIS TIME, FURTHER TESTING IS NOT INDICATED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE SAMPLE QUALITY- FIBRIN. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BD QUALITY WILL CONTINUE TO MONITOR SAMPLE QUALITY COMPLAINTS.

Additional Manufacturer Narrative · 0

THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL LOT NUMBER IS AS FOLLOWS: D4: MEDICAL DEVICE LOT #: 4012740. D4: MEDICAL DEVICE EXPIRATION DATE: 30-JUN-2025. H4: DEVICE MANUFACTURE DATE: 12-JAN-2024. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED AFTER CENTRIFUGATION USING BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES 15 TUBES CONTAINED FIBRIN. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED AFTER CENTRIFUGATION USING BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES 15 TUBES CONTAINED FIBRIN. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1860485 BD VACUTAINER® LH PST¿ II PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 3338012 30382903673767

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown