FDA Adverse Event Malfunction Summary report: N

3005075853-2011-01549

MDR report key: 2061638 · Received April 18, 2011

Report

Report Number
3005075853-2011-01549
Event Type
Malfunction
Date Received
April 18, 2011
Date of Event
March 9, 2011
Report Date
March 23, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). JAWS. THE DEVICE WAS RETURNED FOR ANALYSIS AND UPON INSPECTION THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. IT IS KNOWN FROM THE HISTORY OF THE DEVICE THAT THE CONDITION OF THE JAWS MAY LEAD THAT THE CLIPS DID NOT FORMED AS INTENDED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION WAS REQUESTED.