FDA Adverse Event Malfunction Summary report: N

ENPULSE

MDR report key: 2061625 · Received April 18, 2011

Report

Report Number
2647346-2011-00563
Event Type
Malfunction
Date Received
April 18, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S032
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. IT WAS NOTED THAT THE 75,000 ATRIAL HIGH RATE EPISODES HAD USED UP DEVICE MEMORY AS THE MEMORY HAD NOT BEEN CLEARED AS THE PATIENT HAD NOT BEEN FOLLOWED FOR GREATER THAN ONE YEAR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL HIGH RATES WERE POSSIBLY THE RESULT OF FAR FIELD R WAVE OVERSENSING. THE DEVICE IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENPULSE IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. E2DR01 ASKU

Patients

Seq Age Sex Outcome Treatment
1 12 YR Other (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD