FDA Adverse Event
Malfunction
Summary report: N
BV LIBRA 9" SURVEY
MDR report key: 2061553
·
Received March 16, 2011
Report
- Report Number
- 3003768277-2011-00256
- Event Type
- Malfunction
- Date Received
- March 16, 2011
- Report Date
- February 18, 2011
- Manufacturer
- PHILIPS HEALTHCARE
- Product Code
- IZL
- PMA / PMN Number
- K010762
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE FIELD SERVICE ENGINEER TROUBLESHOOT THE SYSTEM AND FOUND THAT THE POWER DISTRIBUTION UNIT WAS NOT SUPPLYING POWER. HE REPLACED THE "PART NUMBER (B)(4) - ASSY POWER SUPPLY", WHICH SOLVED THE PROBLEM.
Description of Event or Problem · 1
THE CUSTOMER STATES THAT: "THE STAND SHUTS OFF ONCE IN A WHILE".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BV LIBRA 9" SURVEY | IZL | PHILIPS HEALTHCARE | 718030 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |