FDA Adverse Event Malfunction Summary report: N

BV LIBRA 9" SURVEY

MDR report key: 2061553 · Received March 16, 2011

Report

Report Number
3003768277-2011-00256
Event Type
Malfunction
Date Received
March 16, 2011
Report Date
February 18, 2011
Manufacturer
PHILIPS HEALTHCARE
Product Code
IZL
PMA / PMN Number
K010762
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE FIELD SERVICE ENGINEER TROUBLESHOOT THE SYSTEM AND FOUND THAT THE POWER DISTRIBUTION UNIT WAS NOT SUPPLYING POWER. HE REPLACED THE "PART NUMBER (B)(4) - ASSY POWER SUPPLY", WHICH SOLVED THE PROBLEM.

Description of Event or Problem · 1

THE CUSTOMER STATES THAT: "THE STAND SHUTS OFF ONCE IN A WHILE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BV LIBRA 9" SURVEY IZL PHILIPS HEALTHCARE 718030

Patients

Seq Age Sex Outcome Treatment