2-L HUB CONNECT ASSY REP LACEMENT SET
Report
- Report Number
- 1036844-2011-00092
- Event Type
- Malfunction
- Date Received
- March 16, 2011
- Date of Event
- March 7, 2011
- Report Date
- March 11, 2011
- Manufacturer
- ARROW INTL., INC.
- Product Code
- NFK
- PMA / PMN Number
- K020430
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SF
- Reporter Occupation
- UNKNOWN
Narratives
(B)(4). F/U REPORT WILL BE FILED IF ADDITIONAL INFO BECOMES AVAILABLE.
IT WAS REPORTED THAT AN (B)(6) YEAR OLD, VERY EDUCATED LADY HAS HAD THE CURRENT CANNON INSERTED SINCE (B)(6) 2006. SINCE (B)(6) 2010, THE EXTENSION LINES HAD TO BE REPLACED FOUR TIMES IN THE DIALYSIS UNIT, OF WHICH THE LAST THREE OCCURRED (B)(6) 2010, (B)(6) 2011 AND AGAIN ON (B)(6) 2011 ALL BECAUSE OF A CRACKED HUB ON THE EXTENSION LINE. CATHETER CARE GUIDELINES HAD BEEN PROVIDED IN (B)(6) 2010. INVESTIGATION ON THE (B)(6) OCCASION INCLUDED QUESTIONS WITH REGARD TO THE CLEANING PROTOCOL AND CLEANING SOLUTIONS USED, ETC. THE SALES CONSULTANT WAS ASSURED, THEY DID NOT USE ALCOHOL AND/OR ALCOHOL BASED CLEANING SOLUTIONS WITH THIS PATIENT'S (PT'S) CATHETER. THE PT WAS ALSO QUESTIONED WITH REGARD TO HER ROUTINE AT HOME AND ASSURED THE INSERTION SITE AND WHOLE LENGTH OF EXTERNAL CATHETER AND HUBS WERE CLOSED WITH WATERPROOF DRESSING BETWEEN TREATMENTS. HOWEVER, ON INVESTIGATION THIS TIME, IT WAS DISCOVERED THAT THE CAPS ARE SOAKED ROUTINELY IN ALCOHOL WHILE THE PT IS BEING TREATED AND THAT THE CLEANING SOLUTION USED IN THE UNIT IS 70% ALCOHOL BASED. THERE ARE THREE PTS IN TOTAL WITH A CANNON CATHETER IN SITU BEING TREATED AT THIS SPECIFIC UNIT AND NO PROBLEMS WERE EXPERIENCED WITH ANY OF THE OTHER CATHETERS. THE SAME CLEANING PROTOCOLS ARE USED. THE CATHETER CARE GUIDELINES ALSO STATE THAT THE PRODUCT IS ALCOHOL COMPATIBLE HOWEVER, IN THIS INSTANCE WE HAVE INSISTED NO ALCOHOL OR ALCOHOL BASED CLEANING SOLUTIONS BE USED. AS A RESULT, THE PHYSICIAN AND PT HAVE REQUESTED FURTHER INVESTIGATION. THE SALES CONSULTANT HAS SUGGESTED THAT THEY CONSIDER USING HIBITANE IN WATER EXCLUSIVELY AS THE CLEANING SOLUTION IN THEIR UNIT TO PREVENT FUTURE COMPLICATIONS TO THE CANNON CATHETERS THAT MAY BE RELATED TO THE USE OF ALCOHOL AND ALCOHOL BASED SOLUTIONS. THE CONSULTANT ALSO ADDRESSED THE ISSUE WITH THE USERS REGARDING THE ISSUE OF OVER TIGHTENING THE LINES AND USING INSTRUMENTS TO LOOSEN IT ON REMOVAL. THE PT WAS ABLE TO HAVE DIALYSIS WITH NO REPORTED COMPLICATIONS. REFERENCE MDR #1036844-2011-00088 FOR THE FIRST EVENT, 1036844-2011-00089 FOR THE SECOND EVENT, 1036844-2011-00090 FOR THE THIRD EVENT AND 1036844-2011-00091 FOR THE FOURTH EVENT INVOLVING THE SAME PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 2-L HUB CONNECT ASSY REP LACEMENT SET | CHRONIC HEMODIALYSIS CATHETER PRODUCTS | NFK | ARROW INTL., INC. | RF9096856 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR |