HEARTMATE 14V POWER MODULE PT CABLE

MDR report key: 2061493 · Received March 16, 2011

Report

Report Number
2916596-2011-00114
Event Type
Malfunction
Date Received
March 16, 2011
Date of Event
January 17, 2011
Report Date
February 18, 2011
Manufacturer
THORATEC CORP
Product Code
DSQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE 14 VOLT POWER MODULE PT CABLE WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS AND THE REPORTED EVENT OF CONTINUOUS ALARMS WITH A RED HEART ALARM WAS CONFIRMED. A CONTINUITY TEST WAS PERFORMED AND REVEALED THAT THERE WERE OPEN INNER CONDUCTORS ON THE WHITE POWER CABLE. UPON FURTHER EVALUATION, IT WAS NOTED THAT THE WIRE INSULATION WAS STRETCHED AND ELONGATED AND THE INNER CONDUCTORS OF ALL THE WIRES WERE FOUND DAMAGED. THE DAMAGE TO THE WIRES APPEARED TO BE MECHANICALLY INDUCED, AND APPEARED TO BE FROM TWISTING OR PULLING. THE DAMAGE OBSERVED WOULD HAVE RESULTED IN THE REPORTED RED HEART ALARMS. NO FURTHER INFORMATION IS AVAILABLE. THE MANUFACTURER IS CLOSING ITS FILE ON THIS EVENT.

Description of Event or Problem · 1

THE PT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). IT WAS REPORTED BY THE VAD COORDINATOR THAT WHILE THE PT WAS SWITCHING FROM BATTERIES TO THE POWER MODULE/POWER MODULE PT CABLE AT HOME, THE PT CONNECTED THE WHITE POWER CABLE FIRST, AND THEN, WHEN SHE DISCONNECTED THE BLACK POWER CABLE FROM THE BATTERY, SHE HEARD A CONTINUOUS TONE WITH A RED HEART ALARM. THE PT QUICKLY RECONNECTED TO BATTERIES AND ALARM WENT AWAY. THE PT'S DAUGHTER WAS PRESENT AND ATTEMPTED TO SWITCH TO THE POWER MODULE TWICE WITH THE SAME RESULTS. THE POWER MODULE PT CABLE WAS EXCHANGED AND THE ALARMS WERE RESOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTMATE 14V POWER MODULE PT CABLE POWER MODULE PT CABLE (14V) DSQ THORATEC CORP 103426 34756240210

Patients

Seq Age Sex Outcome Treatment
1 42 YR Other