FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2061436 · Received April 18, 2011

Report

Report Number
1423500-2011-04650
Event Type
Malfunction
Date Received
April 18, 2011
Date of Event
March 24, 2011
Report Date
March 24, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS NOT KNOWN BY THE PATIENT, THEREFORE DEVICE EVALUATION CANNOT BE PERFORMED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON THE COMPLETION OF BAXTER'S INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) DURING DWELL 3 WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE POTENTIAL USE ERROR IN THIS COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).

Description of Event or Problem · 1

A PATIENT CONTACTED GLOBAL TECHNICAL SERVICES (GTS) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 WHILE USING THE HOMECHOICE (HC) DURING DWELL 3. GTS HAD THE PATIENT CYCLE POWER TO CLEAR THE ERROR. THE PATIENT ELECTED TO DISCONNECT AND COMPLETE THERAPY WITH A MANUAL DRAIN. PRODUCT SURVEILLANCE CONTACTED THE PATIENT AND CAREGIVER. THE CAREGIVER STATED THE PATIENT DID NOT DISCONNECT DURING THERAPY. THE CAREGIVER STATED THAT THE TUBING GOING INTO THE SUPPLY BAG WAS ABOVE THE CYCLER, CAUSING PROBLEMS AND THE PATIENT BAG DID NOT DISCONNECT. THE CAREGIVER STATED THE SUPPLIES WERE DISCARDED AND THE LOT INFORMATION WAS NOT RETAINED. THE PATIENT HAD NOT NOTIFIED HIS NURSE, BUT WOULD LET HER KNOW. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 67 YR HOMECHOICE