FDA Adverse Event Malfunction Summary report: N

COBAS E411 DISK SYSTEM

MDR report key: 2061423 · Received April 18, 2011

Report

Report Number
1823260-2011-02094
Event Type
Malfunction
Date Received
April 18, 2011
Date of Event
March 9, 2011
Report Date
July 6, 2011
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JLW
PMA / PMN Number
K961481
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO ROOT CAUSE COULD BE DETERMINED. THE MOST LIKELY CAUSE OF THE FALSE LOW TSH VALUE AFTER DILUTION WAS EITHER AN EMPTY DILUENT BOTTLE OR BY INSUFFICIENT SAMPLE VOLUME CAUSED BY CLOT OR FIBRIN PRESENT IN THE UNDILUTED SAMPLE.

Description of Event or Problem · 1

THE CUSTOMER STATED THEY WERE GETTING QUESTIONABLE RESULTS FOR THYROTROPIN (TSH). ON (B)(6) 2011, THE CUSTOMER PERFORMED THREE TESTS ON THEIR COBAS E411 AND RECEIVED RESULTS OF 100UIU/ML ACCOMPANIED BY A DATA FLAG EACH TIME. THE ANALYZER PERFORMED A 1:10 DILUTION AND GENERATED A RESULT OF 0.312 UIU/ML WHICH WAS REPORTED OUTSIDE THE LABORATORY. ON (B)(6) 2011, THE CUSTOMER RAN THREE TESTS WITH RESULTS OF 100 UIU/ML ACCOMPANIED BY A DATA FLAG. THE CUSTOMER RAN A FOURTH TEST AND RECEIVED 0.188 UIU/ML. THE CUSTOMER DID A MANUAL 1:2 DILUTION AND RECEIVED 141.1 UIU/ML WHICH WAS REPORTED OUTSIDE THE LABORATORY. ALL TESTS WERE PERFORMED ON THE SAME ANALYZER. THE CUSTOMER STATED THAT TREATMENT WAS WITHHELD BECAUSE OF THE LOW TEST RESULT FROM (B)(6) 2011, BUT THERE IS NO ALLEGATION OF ANY ADVERSE EVENT. THE LOT NUMBER OF THE TSH IS 16037401. THE FIELD SERVICE REPRESENTATIVE BELIEVES THE CAUSE WAS THE REAGENTS. THE TSH AND DILUENT PACKS WERE BOTH EMPTY. HE REPLACED BOTH. HE CHECKED THE LIQUID LEVEL DETECTOR AND PIPETTOR. CALIBRATION AND QUALITY CONTROL WAS PERFORMED WITH PASSING RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS E411 DISK SYSTEM IMMUNOCHEMISTRY ANALYZER JLW ROCHE DIAGNOSTICS NA NA

Patients

Seq Age Sex Outcome Treatment
1