FDA Adverse Event Malfunction Summary report: N

ETS FLEX ARTICNG

MDR report key: 2061417 · Received April 18, 2011

Report

Report Number
3005075853-2011-01540
Event Type
Malfunction
Date Received
April 18, 2011
Date of Event
March 22, 2011
Report Date
March 25, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INSTRUMENT B: BATCH #: F5UV0L, EXP DATE: 03/23/2014; MFR DATE: 04/23/2009. THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE (A) WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH THREE RELOADS PRESENT. THE RETURNED RELOADS WERE PARTIALLY FIRED WHICH INDICATES THAT THE DEVICE'S FIRING CYCLE WAS INTERRUPTED. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT, AND ALL THE STAPLES FORMED AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT DESCRIBED COULD NOT BE CONFIRMED AS THE CARTRIDGE WAS LOADED INTO THE DEVICE WITHOUT DIFFICULTIES AND DID NOT FALL OUT DURING TESTING. THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE (B) WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS THE CARTRIDGE WAS LOADED INTO THE DEVICE WITHOUT DIFFICULTIES AND DID NOT FALL OUT DURING THE VISUAL AND FUNCTIONAL TESTING.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY PROCEDURE, THE FIRST DEVICE, THE STAPLES WERE LOOSE AND DID NOT STAPLE ALL THE WAY ACROSS. THE SECOND DEVICE, THE RELOAD FELL OFF INSIDE THE PATIENT. THE SURGEON HAD TO RETRIEVE IT. THE TWO WHITE CARTRIDGES AND ONE RELOAD MISFIRED. THEY COMPLETED THE CASE WITH ANOTHER DEVICE. THE SURGERY WAS DELAYED BY TEN MINUTES WHILE THE DEVICE WAS RETRIEVED FROM ANOTHER BUILDING. THERE WAS NO PATIENT CONSEQUENCE. ADDITIONAL FOLLOW-UP: DEVICE DID NOT CUT THE TISSUE AT ALL. SOME STAPLES DEPLOYED SOME WERE STRAIGHT, THE OTHERS WERE FORMED PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ETS FLEX ARTICNG STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK F4NL8F

Patients

Seq Age Sex Outcome Treatment
1 TR45W, 6R45B