FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 2061397
·
Received March 18, 2011
Report
- Report Number
- 3015876-2011-00253
- Event Type
- Malfunction
- Date Received
- March 18, 2011
- Date of Event
- February 17, 2011
- Report Date
- February 17, 2011
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND WAS UNABLE TO VERIFY OR DUPLICATE THE REPORTED FAILURE. PHYSIO-CONTROL OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS THEN RETURNED TO THE CUSTOMER FOR USE. A CONCLUSIVE CAUSE OF THE REPORTED PROBLEM COULD NOT BE DETERMINED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE FAILED TO DETECT A QUIK-COMBO THERAPY CABLE DURING TESTING. THERE WAS NO PT USE ASSOCIATED WITH THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 20 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | 20 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |