FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2061197 · Received April 4, 2011

Report

Report Number
2531779-2011-02091
Event Type
Injury
Date Received
April 4, 2011
Date of Event
February 28, 2011
Report Date
February 28, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. THE EVALUATION OF THE PRODUCT HAS NOT BEEN COMPLETED; THEREFORE, NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

ON (B)(6) 2011, THE PATIENT CONTACTED ANIMAS AND REPORTED ELEVATED BLOOD GLUCOSE (BG) LEVELS. THE PATIENT ALSO CALLED FOR A PUMP REPLACEMENT PER HER DOCTOR'S REQUEST. THE PATIENT INDICATED THAT FOR THE PREVIOUS 3 MONTHS, HER BG LEVELS HAD BEEN ERRATIC. IN SOME CASES, THE PATIENT CLAIMED THAT DEVELOPED KETONES AND SYMPTOMS OF FREQUENT URINATION AND THIRST WHEN HER BG LEVELS WERE ELEVATED. THE PATIENT CLAIMED THAT HER BG LEVELS WOULD RESPOND TO CORRECTIONS VIA INJECTIONS. SHE CLAIMED, HOWEVER, THAT HER BG LEVELS WOULD REMAIN ELEVATED WHEN BOLUSING VIA THE PUMP. NO ISSUES WERE NOTED WITH THE INFUSION SET OR CANNULA(S) AT THE TIME OF ANY BG ELEVATIONS. SHE DENIED THAT THE CANNULAS WERE BENT OR KINKED. SHE ALSO DENIED THAT THERE WERE ANY ISSUES WITH THE CARTRIDGES OR INFUSION SETS. THE PATIENT'S INSULIN WAS NOT DISCOLORED, CLOUDY, OR EXPIRED. THE PUMP'S ALARM HISTORY WAS NEGATIVE. THE ANIMAS REPRESENTATIVE INVOLVED IN THIS CONTACT INFORMED THE PATIENT THAT THERE WAS NO EVIDENCE TO INDICATE A PUMP MALFUNCTION. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PATIENT CLAIMED THAT SHE DEVELOPED SYMPTOMS SUGGESTIVE OF SEVERE HYPERGLYCEMIA. SHE ALSO ALLEGED THAT HER ELEVATED BG LEVELS WERE NOT RESPONDING TO BOLUSES ADMINISTERED VIA THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1 35 YR Life Threatening