FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 2061135 · Received April 18, 2011

Report

Report Number
1423500-2011-04633
Event Type
Malfunction
Date Received
April 18, 2011
Date of Event
March 9, 2011
Report Date
April 13, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE HOMECHOICE PASSED THE RITE ELECTRICAL TEST AND FAILED THE RITE FUNCTIONAL TEST FOR A DISPLAY ISSUE. THE ASSIGNABLE CAUSE WAS THE DIGITAL PRINTED CIRCUIT BOARD (PCB) ASSEMBLY, WHICH WAS SCRAPPED. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN - TIDAL ULTRAFILTRATION REMOVAL SET TOO LOW. A LABEL REVIEW WAS PERFORMED AND FOUND TO BE ADEQUATE FOR THE USE ERROR IDENTIFIED IN THIS REPORT. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION OF IIPV. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH ((B)(4)).

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE, WHICH OCCURRED DURING DRAIN CYCLE 5. THE PROGRAMMED FILL VOLUME WAS 2100 ML AND ULTRAFILTRATION WAS 1783 ML. THIS MEETS BAXTER'S IIPV CRITERIA. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 36 YR