FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2061134 · Received April 18, 2011

Report

Report Number
6000001-2011-02931
Event Type
Malfunction
Date Received
April 18, 2011
Date of Event
March 28, 2011
Report Date
March 28, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K884505
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. VISUAL EXAMINATION CONFIRMED THE REPORTED CONDITION OF A RUPTURED RESERVOIR. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT ONE (1) CE INFUSOR LV 1.5 EXPERIENCED A RUPTURED RESERVOIR DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT AND, THEREFORE, NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.