FDA Adverse Event Malfunction Summary report: N

ARCHITECT C4000 ANALYZER

MDR report key: 2061107 · Received April 18, 2011

Report

Report Number
1628664-2011-00168
Event Type
Malfunction
Date Received
April 18, 2011
Report Date
April 13, 2011
Manufacturer
ABBOTT MANUFACTURING, INC.
Product Code
JJE
PMA / PMN Number
K980367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) CONCOMITANT MEDICAL PRODUCTS: ARCHITECT ICT PRE-AMP BOARD PHOTO COUPLER, LOTS 0834 AND 0840. (B)(4) NEW (B)(4) PROCESS IMPACTED THE LONG TERM RELIABILITY OF THE PHOTO COUPLERS WHICH DEGRADE SLOWLY OVER TIME. THE CAUSE OF THE ARCHITECT ICT PRE-AMP BOARD PHOTO COUPLER DEGRADATION ISSUE WAS DUE TO A NEW MANUFACTURING (B)(4) PROCESS IMPLEMENTED BY THE SUPPLIER. ABBOTT ISSUED A PRODUCT CORRECTION LETTER TO ALL CUSTOMERS WITH AFFECTED INSTRUMENT SERIAL NUMBERS AND IMPLEMENTED A MANDATORY TECHNICAL SERVICE BULLETIN TO INSPECT THE ARCHITECT ICT PRE AMP BOARDS AND REPLACE IF NEEDED.

Description of Event or Problem · 1

ABBOTT FIELD SERVICE INSPECTED THE ARCHITECT C4000 ANALYZER PER MANDATORY TECHNICAL SERVICE BULLETIN (B)(4) AND REPLACED THE ARCHITECT ICT PRE-AMP BOARD AS REQUIRED. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT C4000 ANALYZER AUTOMATED CHEMISTRY ANALYZER JJE ABBOTT MANUFACTURING, INC. C4000

Patients

Seq Age Sex Outcome Treatment
1 ARCHITECT ICT PRE-AMP BOARD PHOTO COUPLER