Narratives

Description of Event or Problem · 1

DATE NOTIFIED: (B)(6) 2010. A MODEL 321 PORTABLE ASPIRATOR (S/N (B)(4)) FAILED TO OPERATE. AN INSPECTION OF THE DEVICE REVEALED A DEFECTIVE BATTERY PACK. THE BATTERY PACK WAS REPLACED, THE DEVICE WAS TESTED TO SPECIFICATIONS AND RETURNED TO THE CUSTOMER. NO DEATH OR INJURY RESULTED FROM THE DEVICE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMPACT PORTABLE ASPIRATOR BTA IMPACT INSTRUMENTATION, INC 321 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other