FDA Adverse Event Other Summary report: N

IMPACT

MDR report key: 2061052 · Received February 22, 2011

Report

Report Number
2242630-2010-00087
Event Type
Other
Date Received
February 22, 2011
Report Date
December 11, 2010
Manufacturer
IMPACT INSTRUMENTATION, INC
Product Code
BTA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DATE NOTIFIED: (B)(6) 2010. A MODEL 305 PORTABLE ASPIRATOR (S/N (B)(4)) FAILED TO OPERATE. AN INSPECTION OF THE DEVICE REVEALED A DEFECTIVE CIRCUIT BREAKER. THE CIRCUIT BREAKER WAS REPLACED, THE DEVICE WAS TESTED TO SPECIFICATIONS AND RETURNED TO THE CUSTOMER. NO DEATH OR INJURY RESULTED FROM THE DEVICE MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMPACT PORTABLE ASPIRATOR BTA IMPACT INSTRUMENTATION, INC 305

Patients

Seq Age Sex Outcome Treatment
1 NA Other