VITALITY 2
Report
- Report Number
- 2124215-2011-06445
- Event Type
- Malfunction
- Date Received
- April 18, 2011
- Date of Event
- January 12, 2011
- Report Date
- January 20, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS DEVICE'S PROGRAMMED PARAMETER SETTINGS AND THERAPY HISTORY WERE USED TO ESTIMATE EXPECTED BATTERY USE TO DATE, THEN COMPARED TO ACTUAL USE. OUR INITIAL ANALYSIS SHOWED THIS DEVICE DID NOT MEET LONGEVITY EXPECTATIONS PER PROGRAMMED VALUES. DURING SUBSEQUENT ANALYSIS, PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. THERE WERE NO KNOWN ALLEGATIONS AGAINST THE DEVICE, AND NO REPORTS OF ADVERSE PATIENT EFFECTS. THIS DEVICE IS INCLUDED IN THE (B)(4) 2007 SHORTENED REPLACEMENT WINDOW ADVISORY POPULATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T167 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | (B)(4)| (B)(4)| (B)(4) |