R SERIES
Report
- Report Number
- 1220908-2024-03966
- Event Type
- Malfunction
- Date Received
- November 5, 2024
- Report Date
- October 16, 2024
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- P160022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE DEVICE WAS RETURNED TO ZOLL ITALY FOR EVALUATION. THE CUSTOMER'S REPORT WAS VERIFIED AND ATTRIBUTED TO A FAULTY PRIMARY MFE SHIELD, SECONDARY MFE SHIELD, PRIMARY ECG SHIELD, AND SECONDARY ECG SHIELD. THE SHIELDS WILL BE REPLACED TO REMEDY THE REPORT. THE DEVICE WILL BE RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
JUSTIFICATION FOR NO UDI, THIS DEVICE WAS MANUFACTURED BUT NOT DOMESTICALLY DISTRIBUTED; IT IS ONLY DISTRIBUTED IN THE INTENDED GEOGRAPHY. THERE IS NO EXISTING UDI REGULATION. ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE DEVICE FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TESTING. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 72268 | R SERIES | DEFIBRILLATOR/PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | 30001001001110112 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |