FDA Adverse Event Malfunction Summary report: N

R SERIES

MDR report key: 20609508 · Received November 5, 2024

Report

Report Number
1220908-2024-03966
Event Type
Malfunction
Date Received
November 5, 2024
Report Date
October 16, 2024
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
P160022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO ZOLL ITALY FOR EVALUATION. THE CUSTOMER'S REPORT WAS VERIFIED AND ATTRIBUTED TO A FAULTY PRIMARY MFE SHIELD, SECONDARY MFE SHIELD, PRIMARY ECG SHIELD, AND SECONDARY ECG SHIELD. THE SHIELDS WILL BE REPLACED TO REMEDY THE REPORT. THE DEVICE WILL BE RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 0

JUSTIFICATION FOR NO UDI, THIS DEVICE WAS MANUFACTURED BUT NOT DOMESTICALLY DISTRIBUTED; IT IS ONLY DISTRIBUTED IN THE INTENDED GEOGRAPHY. THERE IS NO EXISTING UDI REGULATION. ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE DEVICE FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED LEAKAGE CURRENT TESTING. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
72268 R SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION 30001001001110112 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown