FDA Adverse Event Malfunction Summary report: N

SEEDNET CRYOABLATION SYSTEM

MDR report key: 2060852 · Received March 11, 2011

Report

Report Number
9616793-2011-00004
Event Type
Malfunction
Date Received
March 11, 2011
Date of Event
February 10, 2011
Report Date
February 11, 2011
Manufacturer
GALIL MEDICAL LTD.
Product Code
GEH
PMA / PMN Number
K011950
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A PREVENTIVE MAINTENANCE WAS PERFORMED ON THE SYSTEM AND THE SYSTEM WAS FOUND TO BE IN PROPER WORKING ORDER. THE THERMAL SENSOR THAT WAS PROVIDING ERRATIC READINGS WAS NOT RETURNED AND, THEREFORE, COULD NOT BE INVESTIGATED. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. THE PHYSICIAN DID NOT FEEL THAT THE ICE GOT COLD ENOUGH. IT IS UNK IF THE PHYSICIAN WILL RETREAT THE PT.

Description of Event or Problem · 1

A BIOMEDICAL TECHNICIAN AT (B)(6) HOSPITAL CALLED IN AN ISSUE. HE INDICATED THAT THEY CHANGED THE PROBES MULTIPLE TIMES AND THE NEEDLES DID NOT GET COLD ENOUGH. THE MOBILE PROVIDER STATED THAT IN HIS OPINION: THE SYSTEM WORKED FINE. PASSED TEST, MADE VISIBLE ICE ON ULTRASOUND. THERMAL SENSORS DID NOT SHOW COLD TEMPERATURES AS THE DOCTOR WOULD LIKE TO SEE BECAUSE (IN THE MOBILE PROVIDER'S OPINION), DOCTOR DID NOT FREEZE LONG ENOUGH AND PLACED THERMAL SENSORS MAY HAVE BEEN TOO FAR AWAY FROM PROSTATE, ALTHOUGH ONE THERMAL SENSOR DID GIVE ERRATIC TEMPERATURE READINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SEEDNET CRYOABLATION SYSTEM CRYOSURGICAL UNIT, ACCESSORIES GEH GALIL MEDICAL LTD. FPRCH2024

Patients

Seq Age Sex Outcome Treatment
1 UNK