FDA Adverse Event Malfunction Summary report: N

LUPINE DRILL BIT

MDR report key: 2060847 · Received March 11, 2011

Report

Report Number
1221934-2011-00115
Event Type
Malfunction
Date Received
March 11, 2011
Date of Event
March 9, 2011
Report Date
March 10, 2011
Manufacturer
DEPUY MITEK
Product Code
HTW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NOTHING IS BEING RETURNED FOR EVALUATION; COMPLAINT DEVICE DISCARDED AT USER FACILITY. NO LOT HAS BEEN IDENTIFIED, WHICH PRECLUDES CONDUCTING A BATCH RECORD REVIEW TO DETERMINE IF THERE WERE ANY INTERNAL PROCESSING ISSUES WHICH WOULD HAVE CONTRIBUTED TO THE NATURE OF THE PRODUCT COMPLAINT. HOWEVER, WE BELIEVE THAT THIS TYPE OF EVENT IS MOST LIKELY TECHNIQUE RELATED, EXTENSIVE LAB TESTING HAS SHOWN THAT UNDER NORMAL CONDITIONS, THE REPORTED EVENT MODE COULD NOT BE DUPLICATED, HOWEVER, UNDER EXTREME CONDITIONS, THAT IS WHEN THE DRILL GUIDE WAS BENT DURING DRILLING WITH A DRILL BIT, THIS CAUSED THE DRILL BIT TO RUB AGAINST INNER SURFACE OF THE BENT DRILL GUIDE CAUSING SOME METAL TO SHAVE OFF OF THE DRILL GUIDE, THE REPORTED CONDITION WAS THEN DUPLICATED. OUTSIDE OF THIS HYPOTHESIS AND CONSIDERATION, WE CANNOT DISCERN ANY OTHER ROOT CAUSE FOR THE REPORTED ISSUE; AT THIS POINT IN TIME NO CORRECTIVE OR FURTHER ACTION IS WARRANTED.

Description of Event or Problem · 1

OUR REP IS REPORTING TO US THAT DURING AN ARTHROSCOPIC SHOULDER PROCEDURE, THE SURGEON OBSERVED METAL SHAVINGS FALLING INTO THE PATIENT'S BODY WITH THE USE OF A LUPINE DRILL BITS. ALL OF THE DEBRIS WAS REMOVED AND THE PROCEDURE WAS CONCLUDED SUCCESSFULLY WITHOUT FURTHER ISSUE OR HARM TO THE PATIENT. THEY REPORTED THAT THE DRILL BIT WAS VERY DULL AND THE SURGEON HAD TO PRESS DOWN WITH SOME FORCE TO GET IT TO DRILL INTO THE BONE. THE COMPLAINT DEVICES WERE DISCARDED AT THE USER FACILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LUPINE DRILL BIT ARTHROSCOPIC INSTRUMENT HTW DEPUY MITEK 211033 UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK