FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 2060834
·
Received March 11, 2011
Report
- Report Number
- 8020893-2011-00109
- Event Type
- Malfunction
- Date Received
- March 11, 2011
- Date of Event
- January 3, 2011
- Report Date
- March 2, 2011
- Manufacturer
- PURITAN BENNETT CORP.
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE SERVICE REPORT SHOWS, THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE NELLCOR PURITAN BENNETT CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE ANALOG INTERFACE PCB. THE UNIT PASSED EXTENDED SELF TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 840 VENTILATOR | VENTILATOR | CBK | PURITAN BENNETT CORP. | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |