ETS-FLEX - ENDOSCOPIC
Report
- Report Number
- 3005075853-2011-01535
- Event Type
- Malfunction
- Date Received
- April 18, 2011
- Date of Event
- March 21, 2011
- Report Date
- March 23, 2011
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. THE ANALYSIS RESULTS FOUND THAT THE ATW35 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED ON THE DEVICE. THE RELOAD WAS RETURNED FULLY LOADED WITH STAPLES AND WAS NOT LOADED CORRECTLY/COMPLETELY ON THE DEVICE. THE CARTRIDGE WAS TAKEN OFF OF THE DEVICE AND PLACED BACK ON AND IT CLICK INTO PLACE AS INTENDED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH THE RETURNED RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS THE RELOAD WAS LOADED INTO THE DEVICE AS INTENDED AND DID NOT FELL OUT DURING THE VISUAL AND FUNCTIONAL TEST.
IT WAS REPORTED THAT DURING A VATS/ LOBECTOMY PROCEDURE, THE CARTRIDGE FELL OUT OF THE DEVICE AND INTO THE PATIENT. THE PROBLEM OCCURRED PRIOR TO PLACING THE DEVICE ON THE VESSEL. THE DEVICE WAS RELOADED BUT THE CARTRIDGE FELL OUT AGAIN. THE CARTRIDGES WERE ABLE TO BE RETRIEVED THROUGH THE EXISTING INCISION. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETS-FLEX - ENDOSCOPIC | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | G4U85E |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |