FDA Adverse Event
Malfunction
Summary report: N
WARMTOUCH COMPACT WARMING UNIT
MDR report key: 2060779
·
Received March 10, 2011
Report
- Report Number
- 2936999-2011-00182
- Event Type
- Malfunction
- Date Received
- March 10, 2011
- Date of Event
- January 20, 2011
- Report Date
- January 20, 2011
- Manufacturer
- COVIDIEN/FORMERLY TYCO
- Product Code
- DWJ
- PMA / PMN Number
- K913016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
IT WAS REPORTED TO COVIDIEN ON (B)(6) 2011, THAT THE CUSTOMER REPORTED, A PATIENT DEVELOPED RED SPOTS OVER THE BACK POST-OPERATIVELY WHERE THE WARMTOUCH WARMING BLANKET WAS IN PLACE. THE UNIT WAS SENT TO THE HOSPITAL BIOMEDICAL ENGINEERING DEPARTMENT AND WAS TESTED. THE DEVICE WAS FOUND TO OVERHEAT TO 56 DEGREES ON MEDIUM AND HIGH SETTINGS FOR 1 MINUTE. THERE WERE NO ADVERSE LONG TERM PROBLEMS AS FAR AS THE DOCTOR IS AWARE. PATIENT HAS BEEN DISCHARGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WARMTOUCH COMPACT WARMING UNIT | CONVECTIVE AIR WARMER | DWJ | COVIDIEN/FORMERLY TYCO | 5200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |