FDA Adverse Event Malfunction Summary report: N

WARMTOUCH COMPACT WARMING UNIT

MDR report key: 2060779 · Received March 10, 2011

Report

Report Number
2936999-2011-00182
Event Type
Malfunction
Date Received
March 10, 2011
Date of Event
January 20, 2011
Report Date
January 20, 2011
Manufacturer
COVIDIEN/FORMERLY TYCO
Product Code
DWJ
PMA / PMN Number
K913016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

IT WAS REPORTED TO COVIDIEN ON (B)(6) 2011, THAT THE CUSTOMER REPORTED, A PATIENT DEVELOPED RED SPOTS OVER THE BACK POST-OPERATIVELY WHERE THE WARMTOUCH WARMING BLANKET WAS IN PLACE. THE UNIT WAS SENT TO THE HOSPITAL BIOMEDICAL ENGINEERING DEPARTMENT AND WAS TESTED. THE DEVICE WAS FOUND TO OVERHEAT TO 56 DEGREES ON MEDIUM AND HIGH SETTINGS FOR 1 MINUTE. THERE WERE NO ADVERSE LONG TERM PROBLEMS AS FAR AS THE DOCTOR IS AWARE. PATIENT HAS BEEN DISCHARGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 WARMTOUCH COMPACT WARMING UNIT CONVECTIVE AIR WARMER DWJ COVIDIEN/FORMERLY TYCO 5200

Patients

Seq Age Sex Outcome Treatment
1 70 YR