FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2060757 · Received April 8, 2011

Report

Report Number
2531779-2011-02207
Event Type
Malfunction
Date Received
April 8, 2011
Date of Event
October 5, 2011
Report Date
March 5, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL BUTTON KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. A FAMILY MEMBER REPORTED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS HAVE BEEN HARDER TO PRESS AND SLOWER TO RESPOND OVER THE PAST 6 MONTHS. HE DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS, AND STATED THAT THE PATIENT WEARS THE PUMP IN A POCKET OR POUCH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 11 YR