FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 2060744 · Received April 8, 2011

Report

Report Number
2531779-2011-02374
Event Type
Malfunction
Date Received
April 8, 2011
Date of Event
March 10, 2011
Report Date
March 10, 2011
Manufacturer
ANIMAS CORP
Product Code
LZG
PMA / PMN Number
K032257
Removal / Correction Number
2531779-02/25/11-001-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. ALTHOUGH THE CARTRIDGE WAS NOT RETURNED THERE IS NO FURTHER INVESTIGATION NECESSARY WITH RESPECT TO THE LEAK ISSUE. CARTRIDGES WITH LOT# B201581 WERE CONFIRMED TO BE DEFECTIVE. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.

Description of Event or Problem · 1

ON (B)(6) 2011, THE REPORTER CONTACTED ANIMAS ON BEHALF OF HER DAUGHTER (THE PT) ALLEGING A LEAKING CARTRIDGE ISSUE. THE REPORTER CLAIMED THAT DURING THE TIME OF CONCERN, THE PT DEVELOPED A BLOOD GLUCOSE (BG) LEVEL IN THE "400'S" MG/DL. HOWEVER, THE REPORTER DENIED THAT THE PT DEVELOPED SYMPTOMS. AT ONE POINT, THE PT REPORTEDLY COMPLAINED ABOUT HAVING A WET CARTRIDGE. BASED ON THE INFO PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE REPORTER CLAIMED THAT THE PT HAD A LEAKING CARTRIDGE ISSUE. THERE IS NO EVIDENCE HOWEVER THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PT DID NOT HAVE ANY BG READINGS OR SYMPTOMS THAT MEET ANIMAS' CRITERIA FOR A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP CARTRIDGE LZG ANIMAS CORP IR 1200/1250/2020/OTP B201581

Patients

Seq Age Sex Outcome Treatment
1 19 YR