FDA Adverse Event Injury Summary report: N

MBT CEM KEEL TIB TRAY SZ3

MDR report key: 2060677 · Received April 6, 2011

Report

Report Number
1818910-2011-05759
Event Type
Injury
Date Received
April 6, 2011
Date of Event
March 8, 2011
Report Date
March 8, 2011
Manufacturer
DEPUY (IRELAND)
Product Code
NJL
PMA / PMN Number
P830055/S095
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT ASSOCIATED WITH THIS REPORTED EVENT WAS NOT RETURNED FOR EXAMINATION. A SEARCH OF THE COMPLAINT DATABASE DID NOT SHOW ANY ADDITIONAL REPORTS AGAINST THE PRODUCT LOT CODE. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS ABOUT THE ROOT CAUSE OF THE REPORTED TIBIAL TRAY LOOSENING. PROVIDED INFORMATION STATED THE PATIENT EXPERIENCED TRAUMA (FELL) IN 2009, AND THE TIBIAL TRAY RECESSED. NO EVIDENCE WAS FOUND SUGGESTING PRODUCT ERROR WAS A CONTRIBUTING FACTOR AND THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Description of Event or Problem · 1

PATIENT REVISED FOR LOOSE TIBIAL TRAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MBT CEM KEEL TIB TRAY SZ3 87 NJL NJL DEPUY (IRELAND) NA 2420566

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention