HENRY SCHEIN ONE STEP+ HCG URINE STRIP TEST
Report
- Report Number
- 2027969-2011-00756
- Event Type
- Malfunction
- Date Received
- April 8, 2011
- Date of Event
- February 10, 2011
- Report Date
- April 8, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- JHI
- PMA / PMN Number
- K993203
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AK, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
CONTROL LOT: 25MIU/ML, HCG URINE CONTROL LOT: HCG110105-01, 100MIU/ML, HCG URINE CONTROL LOT: HCG100513-01, 278.9IU/ML, HCG URINE CONTROL LOT: HCG110124-01. SUMMARY OF RESULTS: THE RETENTION STIRPS MEET QC SPEC. DETAILS AS BELOW: CORRECT POSITIVE RESULTS WERE OBSERVED AT 3 MINUTE READ TIME WHEN TESTED WITH 25MIU/ML HCG URINE CONTROL. (N=5). CORRECT POSITIVE RESULTS WERE OBSERVED AT 3 MINUTE READ TIME WHEN TESTED WITH 100MIU/ML HCG URINE CONTROL. (N=5). CLEARLY POSITIVE RESULTS WERE OBSERVED AT 3 MINUTE READ TIME WHEN TESTED WITH 278.9IU/ML HCG URINE CONTROL. (N=5). CONCLUSION: FROM RETAIN TESTING WITH IN-HOUSE CONTROLS, THE CLIENT'S RESULT WAS NOT REPLICATED AND THE PRODUCT PERFORMED AS EXPECTED MEETING QC SPECS. VERIFIED THE PRODUCT NUMBER, PRODUCT DESCRIPTION, LOT NUMBER AND BATCH RECORD; NO ABNORMAL RESULTS WERE FOUND. CUSTOMER'S OBSERVATION COULD NOT BE REPRODUCED IN-HOUSE WITH CONTROL SAMPLES. NO FALSE NEGATIVE WAS OBSERVED. THIS ISSUE WILL BE TRACKED AND TRENDED. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
CUSTOMER REPORTED POTENTIAL FALSE NEGATIVE URINE HCG RESULTS USING (B)(6) ONE STEP+ HCG URINE STRIP TEST. ONE PT WAS REPORTED TO HAVE CONFLICTING RESULTS WHEN TESTING FOR URINE HCG. THE END USER PROVIDED THE FOLLOWING INFO: THE FIRST TEST WAS NEGATIVE. A SECOND TEST WAS RUN A FEW MINUTES LATER AND THE RESULT WAS ALSO NEGATIVE. NO SERUM TEST WAS PERFORMED. THE FINAL DIAGNOSIS WAS 'PREGNANT' BUT METHOD OF DETERMINING THIS WAS NOT PROVIDED. THE PT'S LAST MENSTRUAL PERIOD WAS (B)(6) 2011.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HENRY SCHEIN ONE STEP+ HCG URINE STRIP TEST | HCG PREGNANCY TEST | JHI | ALERE SAN DIEGO, INC. | FHC-101-KHS25 | HCG9120190 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |