FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2060533 · Received April 18, 2011

Report

Report Number
6000001-2011-02921
Event Type
Malfunction
Date Received
April 18, 2011
Date of Event
February 1, 2011
Report Date
February 22, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THIS PUMP HAS NEVER PREVIOUSLY BEEN SENT TO BAXTER FOR SERVICE AND, THEREFORE, THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS REPORTED PROBLEM FOR THIS PUMP. A DEVICE HISTORY REVIEW WAS ALSO PERFORMED, FINDING THAT NO EXCEPTIONS WERE NOTED DURING THE MANUFACTURING OF THIS DEVICE. BAXTER DISCONTINUED SERVICE AND CEASED ALL DESIGN RELATED SUPPORT ON FLO-GARD 2M8063 (6201) AND 2M8064 (6301) IN THE U.S. REGION AS OF (B)(4) 2010. BAXTER CONTINUES TO SUPPORT THE PRODUCT IN (B)(4) AND WILL DO SO UNTIL PARTS REMAIN AVAILABLE. BAXTER WILL CONTINUE TO COLLECT PRODUCT COMPLAINTS BUT PERIODIC MONITORING/TRENDING AND ANALYSES OF THE U.S. COMPLAINTS WILL NO LONGER BE REQUIRED. HOWEVER, BAXTER WILL CONTINUE TO MONITOR AND REPORT ON DEATH AND SERIOUS INJURY (DSI) EVENTS TO ASSESS THE IMPACT ON PATIENT SAFETY. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE CONDITION OF THE PUMP 2 DOES NOT WORK DUE TO AN AIR IN LINE ALARM WAS CONFIRMED, DUPLICATED, AND WAS FOUND TO BE DUE TO A DEFECTIVE P2 AIR SENSOR ASSEMBLY. THE P2 AIR SENSOR ASSEMBLY WAS REPLACED AND CALIBRATION WAS PERFORMED TO CORRECT THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A FLO-GARD INFUSION PUMP IN WHICH PUMP 2 DOES NOT FUNCTION. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. DURING DEVICE EVALUATION BY BAXTER ON APRIL 2, 2011, IT WAS DETERMINED THAT THE PUMP 2 DOES NOT WORK, DUE TO AN AIR IN LINE ALARM, IMMEDIATELY AT START INFUSION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1