COLLEAGUE SINGLE CHANNEL CXEPUMP V1.7
Report
- Report Number
- 6000001-2011-02920
- Event Type
- Malfunction
- Date Received
- April 18, 2011
- Date of Event
- March 30, 2011
- Report Date
- March 31, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. UPON COMPLETION OF BAXTER EVALUATION, A FOLLOW UP MEDWATCH WILL BE SUBMITTED. THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. (B)(4).
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH SCREEN DISPLAYING "PUMP MAY STOP UNEXPECTEDLY", WAS CONFIRMED IN THE PUMP'S EVENT HISTORY BUT NOT DUPLICATED DURING PRODUCT EVALUATION. THE DEVICE WAS FOUND TO BE WITHIN SPECIFICATIONS. THEREFORE, NO ASSIGNABLE CAUSE WAS PROVIDED DURING PRODUCT EVALUATION AND NO REPAIR WAS NECESSARY FOR THE REPORTED CONDITION. ADDITIONAL INFORMATION: THIS INVOLVED A COLLEAGUE VERSION 1.7 INFUSION PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 8.11.00. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
THIS IS A REPORT FROM BAXTER (B)(4) OF A COLLEAGUE INFUSION PUMP THAT WAS RETURNED TO BAXTER TECHNICAL SERVICES AND THE SCREEN DISPLAYED, "PUMP MAY STOP UNEXPECTEDLY," DURING BIO MED TESTING. THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE SINGLE CHANNEL CXEPUMP V1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |