SYNCHRON UNICEL DXC 600
MDR report key: 2060406
·
Received April 16, 2011
Report
- Report Number
- 2050012-2011-01175
- Date Received
- April 16, 2011
- Date of Event
- March 16, 2011
- Report Date
- March 16, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
WHILE CUSTOMER TECHNICAL SPECIALIST ASSISTED THE CUSTOMER WITH WASH CONCENTRATE ERRORS, CUSTOMER CHECKED THE HYDRO DRAWER AND THE CUSTOMER STATES THAT THE TOP OF THE LID IS CRUSTY WITH LEAKED WASH CONCENTRATE FLUID. A BCI FIELD SERVICE ENGINEER REPLACED THE WASH CANISTER ASSEMBLY.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) TO REPORT FALSE WASH CONCENTRATE LOW ERRORSN. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON UNICEL DXC 600 | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC600 | N/A |