FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 2060404
·
Received April 16, 2011
Report
- Report Number
- 2050012-2011-01179
- Event Type
- Malfunction
- Date Received
- April 16, 2011
- Date of Event
- March 16, 2011
- Report Date
- March 16, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CUSTOMER CLEANED THE CC PROBES TO FIX THE SUPPRESSED CHEMISTRIES PER CUSTOMER TECHNICAL SPECIALIST'S RECOMMENDATION. AFTER CLEANING THE PROBES CUSTOMER STATED THAT THE CC SAMPLE PROBE COLLAR WASH IS DRIPPING WHICH CAUSED CC SIDE CHEMISTRIES TO BE SUPPRESSED. A BCI FIELD SERVICE ENGINEER (FSE) FOUND GEL IN THE WASH COLLAR VACUUM VALVE, CLEANED, AND THIS RESOLVE THE ISSUE.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO CARTRIDGE CHEMISTRY (CC) SAMPLE COLLAR WASH DRIP. NO INJURY OR EXPOSURE TO BIOHAZARDOUS MATERIALS WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600I | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |