UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2060404 · Received April 16, 2011

Report

Report Number
2050012-2011-01179
Event Type
Malfunction
Date Received
April 16, 2011
Date of Event
March 16, 2011
Report Date
March 16, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K060256
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER CLEANED THE CC PROBES TO FIX THE SUPPRESSED CHEMISTRIES PER CUSTOMER TECHNICAL SPECIALIST'S RECOMMENDATION. AFTER CLEANING THE PROBES CUSTOMER STATED THAT THE CC SAMPLE PROBE COLLAR WASH IS DRIPPING WHICH CAUSED CC SIDE CHEMISTRIES TO BE SUPPRESSED. A BCI FIELD SERVICE ENGINEER (FSE) FOUND GEL IN THE WASH COLLAR VACUUM VALVE, CLEANED, AND THIS RESOLVE THE ISSUE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO CARTRIDGE CHEMISTRY (CC) SAMPLE COLLAR WASH DRIP. NO INJURY OR EXPOSURE TO BIOHAZARDOUS MATERIALS WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 600I N/A