CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS
Report
- Report Number
- 2015691-2011-15240
- Event Type
- Injury
- Date Received
- April 15, 2011
- Date of Event
- December 30, 2010
- Report Date
- March 17, 2011
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- DYE
- PMA / PMN Number
- P860057/S001
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED AND THERE WAS NO NONCONFORMANCE FOUND RELATED TO THIS EVENT.
DEHISCENCE. DEVICE NOT RETURNED. ADDITIONAL MANUFACTURER NARRATIVE: UNFORTUNATELY, THE EDWARDS DEVICE WAS NOT RETURNED FOR ANALYSIS. THE DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED AND THIS DEVICE PASSED ALL MANUFACTURING AND STERILIZATION INSPECTIONS WITH NO NONCONFORMANCE. CALCIFICATION IS A WELL RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION.
EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE. IN THIS CASE, IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AFTER AN IMPLANT DURATION OF APPROXIMATELY 11 YEARS. THROUGH FOLLOW-UP WITH THE HEALTH-CARE PROVIDER, IT WAS LEARNED THAT THE DEVICE WAS EXPLANTED DUE TO A PERIVALVULAR LEAK. ACCORDING TO THE OPERATIVE REPORT, THE PATIENT HAD A PREVIOUS AORTIC VALVE OPERATION WHO PRESENTED WITH A PERIVALVULAR LEAK, CONGESTIVE HEART FAILURE, AND RENAL FAILURE. THEREFORE, HE WAS REFERRED FOR AORTIC VALVE REPLACEMENT. UPON INSPECTION, THE PREVIOUS VALVE WAS NOTED TO BE CALCIFIED AND SOMEWHAT DISEASED, BUT THE LEAK WAS PREDOMINANTLY FROM PERIVALVULAR DEHISCENCE FROM THE RIGHT CORONARY CUSP. THE VALVE WAS REMOVED. THE ANNULUS ACTUALLY APPEARED RELATIVELY HEALTHY. ANOTHER EDWARDS BIOPROSTHETIC VALVE WAS IMPLANTED. THE PATIENT TOLERATED THE PROCEDURE WELL. FURTHERMORE, THERE HAVE NOT BEEN ANY ALLEGATIONS OF A PRODUCT MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARPENTIER-EDWARDS PERIMOUNT RSR PERICARDIAL BIOPROSTHESIS | REPLACEMENT HEART VALVE | DYE | EDWARDS LIFESCIENCES | 2800 | 9G0894 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Required Intervention |