HOMECHOICE PRO
Report
- Report Number
- 1423500-2011-04590
- Event Type
- Malfunction
- Date Received
- April 15, 2011
- Date of Event
- February 5, 2011
- Report Date
- March 23, 2011
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K053512
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER FOR EVALUATION; HOWEVER, THE EVALUATION IS NOT COMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF BAXTER'S INVESTIGATION.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB. THE RITE (RETURN INSTRUMENT TEST/EVALUATION) TEST WAS PERFORMED WHEN THE DEVICE WAS RETURNED TO THE BAXTER (B)(4) FACILITY FOR EVALUATION. THE DEVICE PASSED THE HOMECHOICE RITE FUNCTIONAL TEST, BUT FAILED THE HOMECHOICE RITE ELECTRICAL TEST. THE DEVICE WAS DETERMINED TO MEET SPECIFICATIONS RELATIVE TO THE IIPV IDENTIFIED DURING DEVICE LOG REVIEW. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) IDENTIFIED IN THE DEVICE LOG WAS DETERMINED TO BE INSUFFICIENT DRAIN - USE ERROR, TIDAL UF REMOVAL SET TOO LOW. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4). THIS IS REPORT 2 OF 2.
AN INSTANCE OF INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS IDENTIFIED DURING A REVIEW OF THE RETURNED HOMECHOICE (HC) LOG. ON (B)(6) 2011, THE DRAIN VOLUME WAS 3258ML DURING CYCLE 6.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |