HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-04584
- Event Type
- Malfunction
- Date Received
- April 15, 2011
- Date of Event
- March 21, 2011
- Report Date
- March 21, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS NOT KNOWN BY THE PATIENT, THEREFORE, DEVICE EVALUATION CANNOT BE PERFORMED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON THE COMPLETION OF BAXTER'S INVESTIGATION.
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE LOT NUMBER IS UNKNOWN THEREFORE A BATCH REVIEW WAS NOT PERFORMED. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE POTENTIAL USE ERROR IN THIS COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
A PATIENT CONTACTED GLOBAL TECHNICAL SERVICES (GTS) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 WHILE USING THE HOMECHOICE (HC) DURING DWELL 3 OF 5. THE PATIENT HAD THREE BAGS CONNECTED, SOLUTION IN THE HEATER AND FINAL BAGS ONLY. GTS HAD THE PATIENT CYCLE POWER. THE PATIENT ELECTED TO COMPLETE THERAPY MANUALLY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.