FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 2060239 · Received April 15, 2011

Report

Report Number
2954323-2011-02968
Event Type
Malfunction
Date Received
April 15, 2011
Date of Event
April 8, 2011
Report Date
June 2, 2011
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS IS AN INITIAL REPORT. THE CUSTOMER'S PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED.

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE READINGS THE CUSTOMER REPORTED WERE FOUND IN METER MEMORY.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 444 MG/DL AND 121 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.