SYNCHRON® LX20 CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-01102
- Event Type
- Malfunction
- Date Received
- April 15, 2011
- Date of Event
- March 16, 2011
- Report Date
- March 16, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- CGA
- PMA / PMN Number
- K965240
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
SAMPLES WERE SERUM TYPE. CUSTOMER NOTICED THEIR QC HAD DRIFTED LOW AFTER THESE SAMPLES WERE RUN AND REPORTED. CUSTOMER DECLINED SERVICE CALL UNTIL THEY PERFORMED 4 MONTH MAINTENANCE THAT WAS DUE, TAPED A BUBBLE OUT FROM THE FACE OF THE GLUCOSE ELECTRODE, FOUND A LOOSE REAGENT TUBING FITTING AND CORRECTED, RECALIBRATED AND QC WAS NOW IN SPECIFICATIONS. CUSTOMER WAS UP AND RUNNING TO THEIR SATISFACTION WHEN HOTLINE CALLED BACK THE NEXT DAY. ROOT CAUSE OF THIS EVENT IS UNKNOWN.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) AND REPORTED THAT THREE (3) ERRONEOUSLY LOW GLUCOSE (GLUCM) PATIENT RESULTS WERE GENERATED BY THE SYNCHRON LX20 CLINICAL SYSTEM AFTER QC LOW DRIFT OCCURRED. THE RESULTS WERE REPORTED OUT OF THE LAB. CUSTOMER RE-TESTED PATIENT SAMPLES ON A DIFFERENT INSTRUMENT AND AMENDED THE RESULTS. CUSTOMER DID NOT THINK PATIENT TREATMENT WAS AFFECTED BASED ON THE SMALL DIFFERENCE BETWEEN ERRONEOUS AND AMENDED RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® LX20 CLINICAL SYSTEM | CLINICAL CHEMISTRY ANALYZER | CGA | BECKMAN COULTER INC. | LX20, WO CAP PIERCER |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |